The Mako R/F Probe, paired with the Mako Base Unit, offers industry-leading accuracy (±1.5% kV measurement uncertainty) and exceptional sensitivity, covering the full range of dose rates.

Its unique design ensures easy setup in the X-ray beam, flexible connections between the Probe and Base Unit, and a small radiological footprint—perfect for seamless measurements without disrupting the AEC system.

 

Ideal for Radiography, Fluoroscopy, CT, and Dental (Intraoral & Panoramic) applications.

Mako R/F Probe

Quantity
  • Automated, Data-Driven Excellence

    Step into a world where data collection is streamlined and efficient. Ocean Next™ offers automated routines and complete traceability, with real-time data gathering from Mako and its probes. Pre-defined templates ensure smooth collaboration and quick setup, keeping you ahead of the curve.

    The Future-Proof Solution

    Mako is designed to evolve with your needs. Its multi-modal design and continuous innovation mean it will always adapt to include new functions, probes, and expanded applications, ensuring it remains your go-to tool for the future.

    All-in-One Measurement Solution

    Mako is more than just a tool—it's a complete solution. Compact, cable-free, and designed for multiple uses, it handles all the key parameters you need. With its high sensitivity, Mako delivers quick and accurate measurements, providing precise results faster than other tools.

  • General Specifications

    • Size: 122.5 x 28 x 14.5 mm (4.82 x 1.1 x 0.57 in)
    • Weight: 55 g (1.94 oz)
    • Detector active areas:
      • Smallest 0.9 x 0.9 mm²
      • Largest 7 x 7 mm²
    • Detector position check: Orientation independent, no check required
    • Backscatter Protection: Yes

     

    Measuring Specifications (Mako R/F)

    • Tube Potential
      • Range: 35 – 155 kV
      • Min dose rate: 2 µGy/s
      • PPV and kVp: Yes
      • Uncertainty: ±1.5 %
    • Dose
      • Range: 1 nGy – 9999 Gy
      • Uncertainty: ±5 % or 5 nGy
    • Dose Rate
      • Range: 1 nGy/s – 500 mGy/s
      • Resolution: 0.1 nGy/s
      • Trigg level:
        • 25 nGy/s @ exp. time >200 ms
        • 3 µGy/s @ 1 ms
        • 0.3 µGy/s @ 10 ms
      • Uncertainty: ±5 % or 1 nGy/s
    • Time
      • Range 0.33 ms – 48 h
      • Resolution: 0.33 ms Bandwidth Up to 1 kHz
      • Uncertainty: ±0.5 % or 0.33 ms
    • HVL
      • Range: 1 – 15 mm Al
      • Min dose rate: 0.5 µGy/s
      • Uncertainty: ±5 % or 0.2 mm Al
    • Total Filtration
      • Range 1 – 90 mm Al equivalent
      • Min dose rate 2 µGy/s
      • Uncertainty ±10 % or 0.2 mm Al
    • Pulses
      • Range 1 – 5 000 000
      • Min peak dose rate 0.25 µGy/s
      • Min pulse width 0.33 ms
    • Pulse Rate
      • Range 0.2 – 260 Hz
      • Min peak dose rate 0.25 µGy/s
      • Min pulse width 0.33 ms
    • Waveforms
      • Range 0 – 240 s Resolution 0.33 ms at exposures <15 s
      • Bandwidth kV 1 kHz
      • Bandwidth Dose Rate 1 kHz